Description
Main OEM |
Dräger |
OEM Sensor Part No. |
6850645 |
Output Voltage |
(24 – 32 μA) in air at 25°C sea level |
Measurement range |
0 to 100[%] O2 |
Output Connector |
Slip rings, x 3 |
Accuracy |
Within +/- 1[%] of full scale at constant temperature and pressure |
Response time |
Less than 12 seconds to 90[%] of final value |
Operating humidity |
0 to 99[%] RH (non-condensing) |
Operating temperature |
0°C to 40°C (32°F to 104°F) |
Storage temperature |
0°C to 50°C (32°F to 122°F) |
Interference |
Less than 0.5[%] O2 response to: 10.0[%] CO2 balance N2 80.0[%] N2 O balance N2 7.5[%] Halothane balance N2 7.5[%] Isoflurane balance N2 7.5[%] Enflurane balance N2 9.0[%] Sevoflurane balance N2 20.0[%] Desflurane balance N2 |
Expected Life |
33 months in air |
Warranty |
15 months against defects in materials and workmanship |